The impact of different cavity design parameters on intraoral scanner accuracy in endocrown restorations
American Journal of Dentistry, cilt.39, sa.4, ss.211, 2026 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 39 Sayı: 4
- Basım Tarihi: 2026
- Dergi Adı: American Journal of Dentistry
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, MEDLINE
- Sayfa Sayıları: ss.211
- Marmara Üniversitesi Adresli: Evet
Özet
Purpose: To evaluate the effects of different tooth structure thicknesses and cavity depths on the accuracy of intraoral scanners (IOS) in endocrown restorations. Methods: A mandibular typodont model was used, with the left mandibular first molar replaced by an extracted real human tooth. Endocrown cavities with varying depths (3.5 mm and 5 mm) and remaining wall thicknesses (1.5 mm and 0.5 mm) were prepared. Master models were scanned using a high-precision extraoral scanner, and the data were exported in STL format. Ten intraoral scans were performed for each group using the TRIOS 3 IOS. Scanning accuracy was analyzed using the Root Mean Square (RMS) value for trueness and the Interquartile Range (IQR) for precision in Geomagic Design X software. The differences in RMS values among the groups were examined using Independent Samples T test and one-way ANOVA test. Results: Significant differences in RMS and IQR values were observed between groups with different cavity depths and remaining tooth structure thicknesses (P< 0.001). Groups with thinner remaining tooth structures and deeper cavities exhibited greater deviations, indicating reduced scanning accuracy.