A method for obtaining the pure annihilation peak of Na-22 radioisotope


Celiktas C., Tav C.

RADIATION MEASUREMENTS, cilt.39, sa.5, ss.569-572, 2005 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 39 Sayı: 5
  • Basım Tarihi: 2005
  • Doi Numarası: 10.1016/j.radmeas.2004.10.013
  • Dergi Adı: RADIATION MEASUREMENTS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.569-572
  • Anahtar Kelimeler: gamma ray spectroscopy, Na-22 and peak selection, positron annihilation radiation, PULSE-SHAPE DISCRIMINATOR
  • Marmara Üniversitesi Adresli: Evet

Özet

By means of pulse separation technique, the annihilation gamma ray peak, which has great importance for positron annihilation lifetime spectroscopy (PALS), was discriminated from the superimposed photon spectrum generated by Compton effect, backscattering, and background in the 0.5 MeV energy region. Energy resolution of 4.5% was obtained for the discriminated annihilation peak of Na-22 positron emitter radioisotope. (c) 2004 Elsevier Ltd. All rights reserved.